User Guide

20
The LCD will then display the last pH calibration date.
Press DISPLAY to scan remaining data, in the following order:
last calibration time;
electrode offset value in mV ("OFF" appears on the secondary LCD);
electrode slope in mV/pH ("SLP" appears on the secondary LCD);
first calibration point buffer;
second calibration point buffer (only if a 2-point calibration has
been performed).
If calibration was performed with an old or dead probe, the
message "old probe" or "dead probe" will scroll on LCD.
The meter then returns to normal mode. Press (ALT+) CAL DATA
to escape before viewing all the data.
To view the mV calibration data
Press (ALT+) CAL DATA when the meter displays mV reading.
The LCD will then display the last mV calibration date.
Press DISPLAY to scan remaining data, in the following order:
last calibration time;
first calibration point;
second calibration point;
third calibration point (only if a 3-point calibration has been
performed).
The meter then returns to normal mode. Press (ALT+) CAL DATA
to escape before viewing all the data.