• The LCD will then display the last pH calibration date.
• Press DISPLAY to scan remaining data, in the following order:
last calibration time;
electrode offset value in mV ("OFF" appears on the secondary LCD);
electrode slope in mV/pH ("SLP" appears on the secondary LCD);
first calibration point buffer;
second calibration point buffer (only if a 2-point calibration has
been performed).
If calibration was performed with an old or dead probe, the
message "old probe" or "dead probe" will scroll on LCD.
• The meter then returns to normal mode. Press (ALT+) CAL DATA
to escape before viewing all the data.
To view the mV calibration data
• Press (ALT+) CAL DATA when the meter displays mV reading.
• The LCD will then display the last mV calibration date.
• Press DISPLAY to scan remaining data, in the following order:
last calibration time;
first calibration point;
second calibration point;
third calibration point (only if a 3-point calibration has been
performed).
• The meter then returns to normal mode. Press (ALT+) CAL DATA
to escape before viewing all the data.