User Guide

15
PARAMETER SETTINGS
Once the parameter code (and password if
needed) has been entered, the current
value of the selected parameter will be
displayed on the primary LCD and the
parameter code on the secondary one. The parameter or a part of it
will blink (e.g. only the hour blinks if time has been selected).
Enter the new value using the arrow keys.
If there is another part of the parameter to be set (e.g. minutes
for the time), press DISPLAY to gain access and then enter the
new value using the arrow keys.
Press CFM to confirm the value.
If the entered value is not acceptable, the "WRONG" indication
will be displayed for a few seconds and then the meter will ask for
a correct value.
If the confirmed value is accepted, the meter will pass to the next
parameter (asking for the password if it is protected).
Note: If (ALT+) SETUP are pressed before parameter value confir-
mation, the meter will not update the parameter and after
escaping will ask for a new setup code.
The following table lists the setup codes along with the description of the
specific setup items, their valid values and the factory settings (default):
Code Valid values Default
00 Instrument ID code 0000 to 9999 0000
01 Current time
1
hh:mm 00:00
02 Current date
1
dd.mm 01.01
03 Current year
1
YYYY 1998
10 Calibration alarm time-out 01 to 99 days , OFF OFF
20 Auto-Off/Power down time-out ON, OFF ON
21 Firmware version
22 Battery level test
90 HI 504910 terminal mode
99 Password
2
0000 to 9999 0000