LEDs TEST
• To enter the LEDs test procedure select the setup item t.03
and an “OFF” blinking message will appear on the LCD.
• Press the (or ) key once and the message will switch to
a blinking “GO”.
• Press CFM key to confirm or the (or ) key again to
return to the previous status.
• If confirmation is given when the “OFF” message is blink-
ing, no action is performed and it will move to the first
setup item of the group (t.00).
• Once confirmed the test, all the LEDs (if anyone was ac-
tive) are switched off and the message “LEDs test - Press
CFM to escape” will scroll on the primary LCD during all
the test.
• Some keys are used to toggle LEDs ON and OFF:
- the DISPLAY key toggles the green LED;
- the (ALT+) CAL DATA keys toggle the red LED.
• To exit the test press the CFM key and the previous configu-
ration of the LEDs is reestablished. The instrument will remain
in the setup menu and move to the first setup item of the
group (t.00).