CONNECTIONS
The connections for the 2-pin RS485 terminal provided are
as follows:
The instrument has no internal line termination. To terminate
the line, an external resistor equal to the characteristic line
impedance (typically 120Ω) must be added at both ends of
the line.
The RS485 can connect up to 31 Digital Transmitters on the
same physical network. All the units are slave devices and
are monitored and controlled by a single master station (typi-
cally an industrial PLC or PC).
Each unit is identified by its RS485 address, included in the
00 to 99 interval, which corresponds to the Instrument ID
configured through the setup item G.11.
(If the instrument does not recognize the address within the
command string, then it discards all the following bytes).
To avoid erroneous readings in Open-Line
conditions, pull-up and pull-down resis-
tors should be connected as shown (Fail
Safe Open Line protection method).
The Fail-Safe resistors are connected only to one unit in the
line, and their value depends on the application and char-
acteristic impedance of the connection cable.