• If an offset calibration has been
made, the instrument will turn to
“slope” calibration (as indicated by
the “SLO” message on the second-
ary display. The slope value is shown
on the primary LCD and the first digit
is blinking to permit modifications).
• Press the or key to modify the value or key to move
to the next digit.
• Once selected the desired value, press CFM to confirm.
• After confirmation the instrument will turn back to normal
mode.
• If the slope is invalid, the ”WRONG” indicator will blink
on the LCD.
Note Press DISPLAY or (ALT+) CAL DATA to exit calibration. The
slope will be set to the default value (57.5 mV/pH).
Note After direct selection of offset and slope, calibration data can
be viewed by pressing (ALT+) CAL DATA while in normal
mode, but no pH-buffer values will be displayed during the
data scrolling on the LCD.