User Guide
32
LAST CALIBRATION DATA
If the instrument is set as pH meter, the following data about
the last calibration are stored in the EEPROM:
• Date
• time
• offset, in mV
• slope, in mV/pH
• up to two buffers.
If the instrument is set as ORP meter, the data stored in the
EEPROM are the following:
• Date
• time
• first calibration point
• second calibration point.
While displaying these data the HI 504910 continues to
acquire and transmit measurements.
To enter the last calibration data mode,
press (ALT+) CAL DATA.
If the meter has never been calibrated
or an EEPROM reset has occurred, no
calibration data is shown when CAL
DATA is pressed. The “no CAL” mes-
sage will blink for a few seconds, then
the meter skips back to the previous
mode.
If the meter is configured to measure pH, once entered the
last calibration data, the following messages could scroll twice
on the primary LCD before showing the calibration date:
• “Old probe”
• “Dead probe”.
Otherwise the last calibration date will
appear on the primary LCD displayed
as DD.MM format, while the second-
ary display will show the last two digit
of the year.
Pressing the key, the meter will cycle through the following
steps in reverse order, i.e. beginning from last buffer.