User Guide
29
Clean and check the electrode.
• pH electrode broken or leakage (10) : The electrode
glass is broken or short circuits have been caused by moisture
or dirt.
Check pH probe and cable; replace them if necessary.
• Reference electrode broken or dirty (11) : The refer-
ence electrode is soiled or broken. Also the reference elec-
trode or the matching pin could not have been immersed
in the solution.
Clean the electrode and replace it when necessary.
• Old pH probe (12) : The pH probe is old (offset outside
of ±30 mV or slope outside the range 53.5 to 62 mV/pH).
The electrode is still working, but it is recommended to
clean or replace it.
• Dead pH probe (13) : Offset is outside of ±60 mV or
slope outside the range 40 to 70 mV/pH. Readings are not
reliable.
Try to clean the probe; if the error persists after cleaning,
replace the probe immediately.
• No calibration (14) : The device has lost its calibration
due to an EEPROM reset.
Calibrate again all the magnitudes.
• Temperature probe broken (20) : Check the tempera-
ture probe wires and connections.
Replace the probe if necessary.
• Power reset (90) : The device was restarted due to a
power failure.
• EEPROM corruption (91) : The EEPROM data is dam-
aged. Reset the EEPROM, then recalibrate and reconfigure
completely the instrument.
• Watchdog reset (92) : The device was restarted due to a
watchdog reset or a manual reset done through the key-
board.