User Guide

28
MEASURE MODE
The measure mode is the normal mode for the instrument.
During measure mode, pH and temperature or ORP mea-
surements are acquired and sent to HI 504 or to a PC with
HI 92500 or another software implementing the communi-
cation protocol described in this manual (see the “RS485
communication section”).
In a normal situation, during measure mode the green LED is
ON and the red LED is OFF (LEDs are located inside the
instrument case).
When the Setup mode or the Calibration mode are entered
through the HI 504920 (or through the HI 504910 inter-
nal keyboard), both the green and the red LEDs are ON.
The red LED ON warns the user that the HI 504910 is not
measuring and transmitting measurements.
Note In this situation, an HI 504 connected to HI 504910 will
be in hold mode.
In case of an error occurs in HI 504910, the green LED
turns OFF and the red LED flashes.
Note When measurements are in overflow, the internal LCD of the
HI 504910 or that of HI 504920 (in terminal mode) will
show the blinking upper or lower boundary value.
ERRORS
When an error occurs, the green LED turns OFF and an
alarm indication is given by the flashing red LED.
Below are listed all the errors with the correspondent error
code.
The error code is reported in the event log file, which can be
examined by means of the DISPLAY key while in normal mea-
sure mode (see the “Event log file scrolling” section). This is
necessary for identifying which error is active and makes the
red LED flash.
Life check error (03) : The pH reading did not vary more
than ±0.10 pH (for pH) or the mV reading did not vary
more than ±10 mV within the time selected through the
“Life check time” setup item (I.11).