User Guide

26
Code Valid Values Default Present
for ORP
READING OFFSETS (“OFFS”)
F.00 pH or ORP actual value measured value -1.00 pH or measured yes
(see note 7) -200 mV to measured value value
+1.00 pH or +200 mV
F.01 pH or ORP reading offset -1.00 to +1.00 pH or 0.00 pH yes
adjustment -200 to +200 mV or 0 mV
F.10 Temperature actual value measured value -10.0 ºC to meas. value no
(for ATC only, see note 8) measured value +10.0 ºC (see note 7)
F.11 Temperature reading offset -10.0 to +10.0 ºC 0.0 ºC no
adjustment (for ATC only, see note 8)
TEST (“tESt”)
t.00 Display test “OFF”: To skip without testing “OFF” yes
“GO”: To start the display test
t.02 EEPROM test “OFF”: To skip without testing “OFF” yes
“GO”: To start the display test
t.03 LEDs test “OFF”: To skip without testing “OFF” yes
“GO”: To start the display test
Notes
(1): The calibration password allows only calibrations, while
the general password allows everything (including calibra-
tion). Obviously, the general password and the calibration
password cannot be viewed among other items when the
“SETUP” key is pressed without entering the right general
password. The instrument is sold with a “0000” general pass-
word.
(2): When a wrong setup value is confirmed, the instrument
does not skip to the next setup item, but remains in the cur-
rent item displaying a blinking “WRONG” indicator till the
parameter value is changed by the user (the same thing hap-
pens also for the setup code selection).
(3): See the “Temperature compensation” subsection for more
details on how the Automatic temperature compensation and
Manual temperature compensation work.