The below table lists the setup codes along with the description of
the specific setup items, their valid values and whether the item is
present for ORP mode.
Code Valid Values Default Present
for ORP
GENERAL (“GEnE”)
G.00 pH/ORP input “PH”, “OrP” (see note 4) “PH” yes
G.01 Temperature compensation “AtC”: Automatic “AtC” no
“USEr”: Manual (see note 3)
G.02 Manual or probe error -30 to 130.0 ºC 25.0 no
temperature (see note 3)
G.10 Factory ID 0000 to 9999 (see note 9) 0000 yes
G.11 Instrument ID 00 to 99 (see note 9) 00 yes
alias RS485 address
G.98 Calibration password 0000 to 9999 (see notes 1, 9) 0000 yes
G.99 General password 0000 to 9999 (see notes 1, 9) 0000 yes
OUTPUT (“OutP”)
O.30 Baud rate (see notes 9, 12) 1200, 2400, 4800, 9600, 19200 2400 yes
O.31 Modem calls answer enable “OFF”: disabled “OFF” yes
(see note 9) “On”: enabled
INPUT (InPU”)
I.11 Life check time “OFF”: life check disabled “OFF” yes
(see note 5) 1: 1 hour
2: 2 hours
4: 4 hours
I.12 Minimum pH probe slope 45 to 75 mV/pH 45 mV/pH no
I.13 pH electrode impedance “OFF”: disabled “On” no
test enable (see note 10) “On”: enabled
I.14 Reference electrode impedance “OFF”: disabled “On” yes
test enable “On”: enabled
I.15 Max ref. electrode impedance 0.5 to 100.0 kΩ 50.0 kΩ yes
I.17 Mains frequency (see note 11) 50 or 60 Hz 50 Hz yes