User Guide
22
SETUP MODE
The Setup Mode allows the user to set all needed character-
istics of the meter.
Press (ALT+) SETUP and enter the password when
the device is in idle or control mode.
If the correct password is not entered, the user can only view
the setup parameters (except for passwords) without modify-
ing them (and the device continues to acquire and transmit
measurements).
To each setup parameter (or setup item) is assigned a three
characters (one letter followed by two digits) setup code which
is entered and displayed on the secondary LCD. The first
character identifies the group of setup items, while the two
digits identify the particular item within that group.
The setup codes can be selected after password is entered
and CFM key is pressed. When CFM is pressed, the current
setup item is saved on EEPROM and the following item is
displayed.
The possible transitions in setup mode are the following:
ENTERING THE PASSWORD
• Press (ALT+) SETUP to enter the
setup mode. The primary LCD
will display “0000”, while the
secondary LCD shows “PAS”.
The first digit of the primary LCD
will blink.
• Enter the first digit of the pass-
word by using the or key.
• Then move to the next digit with
and enter a digit as described
above. Continue for the last two
digits.
• When the whole password has
been inserted, press CFM to
confirm it.