User Guide

19
• For the second calibration point it is possible to choose
between 350 mV and 1900 mV. Pressing the or key
the value on the secondary LCD will switch between the two
possibilities.
• Once selected the second calibration point, set the HI
931001 (350 mV) or HI 8427 (350 or 1900 mV) simula-
tor to the same value and the calibration proceeds as for
the first point.
Note In ORP calibration there is no time-out.
TEMPERATURE CALIBRATION
Check that the code 90 has been entered in Setup mode.
• Prepare a beaker containing crushed ice and water at 0°C
(32°F) and another one with hot water at 25°C (77°F) or
50°C (122°F).
• Use a Checktemp or another calibrated
thermometer with a resolution of 0.1° as a
reference thermometer.
• Immerse the temperature probe in the bea-
ker with ice and water as near to the
Checktemp as possible.
Note The instrument can support Pt100 or Pt1000 temperature
sensor and calibration can be performed with anyone of these
two probes.
• After entering the calibration mode, move through the menu
(using the or key) to choose the temperature and the
correct kind of used probe; the primary
LCD shows “°C” and the secondary LCD
gives indication about the kind of probe
(“100” indicates a Pt100 probe, while
“1000” stands for a Pt1000 probe).
Press the CFM key to confirm selection.
• The calibration has to be performed over two points: the
first point has to be 0°C and the second one can be cho-
sen between 25°C and 50°C.
Once confirmed the type of calibration,
the primary LCD will display the current
temperature measure and the second-