User Guide
18
Note If (ALT+) SETUP are pressed instead of CFM, the calibration
value selection is aborted and the meter reverts back to a
two-point calibration.
Note If the calibration is performed by means of the HI 504910
internal keyboard and LCD (not through the HI 504920
meter), the calibration date and time are set to a default 1/1/
2002 00:00 : in this case disable the calibration time-
out in HI 504.
ORP CALIBRATION
Check that the code 90 has been entered in Setup mode.
To perform ORP calibration it is necessary to connect an HI
931001 or HI 8427 simulator to the BNC socket.
The meter has to be set as ORP controller.
• A two-point calibration has to be performed: the first point
value is 0 mV and the second one can be chosen between
350 mV and 1900 mV.
• Enter the calibration mode, select the ORP calibration (use
the and keys to move through setup menu) and press
the CFM key.
• Set the HI 931001 or HI 8427 simulator to 0 mV.
• The primary LCD will display the cur-
rent mV measure and the secondary
LCD will show the first calibration
point (0 mV).
• When the reading becomes stable, if the ORP value is
close to the calibration point, the “CFM” indicator starts
blinking; otherwise the “WRONG” indicator blinks and
the “CAL” is fixed on.
• In the first case press CFM to confirm calibration. The
meter will proceed showing the scrolling message “Press
CFM again to confirm the current buffer or right to es-
cape” (to prevent from confirming the calibration point
inadvertently). Pressing again CFM the secondary LCD will
display the second calibration point.
• In the second case (blinking “WRONG”) the meter will
remain in the WRONG state until the reading becomes
unstable or the calibration mode is exited by pressing the
CAL key.