User Guide
99
ing (setup item L.14).
• To exit the test press the CFM key; the instrument will re-
main in the setup menu and move to the next setup item
(t.08).
Note During the test the relays are activated, the primary LCD will always
show the “ACLE” message and on the secondary display will be
shown the actual status of the digital input (“On” or “OFF”).
WATCHDOG TEST
This test executes a dummy dead loop that causes a watch-
dog reset signal to be generated.
• To enter the test procedure select the setup item t.08 and an
“OFF” blinking message will appear on the LCD.
• Press the (or ) key once and the message will switch to
a blinking “GO”.
• Press CFM key to confirm or the (or ) key again to
return to the previous status.
• If confirmation is given when the “OFF” message is blink-
ing, no action is performed and it will move to the beginning
of the test setup menu.
• Once confirmed the test, the “Reset test” message will scroll
once on the primary LCD, and then the start-up procedure
is initiated.