(analog output 1) and O.21 (analog output 2).
• A new output value can be edited manually. It is possible
to change the value of the first blinking digit by pressing
the or key.
• To move to the next digit press the key once; the first digit
will be fixed and the second one will start blinking.
• Pressing repeatedly the key will result in cycling through
the digits on the primary LCD.
• Once a value is selected, the corresponding current is
immediately erogated by the selected output and no confir-
mation is required.
• The minimum and maximum values let out are 3.6, 22 mA
for the 4-20 mA output and 0, 22 mA for the 0-20 mA
output. This depends on the possibility of the fault currents
to be let out (see “Alarm - error configuration” section for
more details).
• To verify the erogated current use a multimeter connected
to the corresponding output.
• To exit the test press the CFM key; the instrument will re-
main in the setup menu and move to the next setup item.
HOLD DIGITAL INPUT TEST
This test is made to verify if the instrument recognizes the
digital input signal at the hold input.
• To enter the test procedure select the setup item t.06 and an
“OFF” blinking message will appear on the LCD.
• Press the (or ) key once and the message will switch to