• During this time the instrument performs the EEPROM check,
and if the checksum is correct, the “Stored data good”
message will scroll on the primary display.
• After that, the meter will remain in setup menu and move to
the next setup item.
• If the checksum fails, a fault alarm is generated and the
following message appears on the LCD: “Stored data error
- Press “UP” button to reset stored data or “RIGHT” button
to ignore”.
• If the key is pressed the process controller restarts opera-
tion, but alarm actions are performed as configured by the
user (see “Alarm - Error configuration” section).
Note that the device will be in Hold mode in any case.
• If the key is pressed, all the data stored in the EEPROM
are erased and the default values loaded.
• Once performed the selected action, the instrument will
remain in the setup menu and move to the next setup item
(t.03).
RELAYS AND LEDS TEST
• To enter the relays and LEDs test procedure select the setup
item t.03 and an “OFF” blinking message will appear on
the LCD.
• Press the (or ) key once and the message will switch to
a blinking “GO”.
• Press CFM key to confirm or the (or ) key again to
return to the previous status.
• If confirmation is given when the “OFF” message is blink-
ing, no action is performed and it will move to the next
setup item (t.04).
• Once confirmed the test, all the relays and LEDs (if anyone
was active) are switched off and the message “Relays and
LEDs test - Press CFM to escape” will scroll on the primary
LCD during all the test.