User Guide
92
If the error is due to impossible communication with the
EEPROM or the RTC, all the pH/ORP controller tasks are
stopped, the alarm relay is de-energized, the red LED blinks
and the “Serial bus error” slides forever (repair can not be
postponed).
A software watchdog is provided in order to detect dead loop
conditions or other causes that make the software stuck. If it
happens, a software reset is generated after a time-out of 1
second.
SELFTEST PROCEDURES
The selftest procedure can be made only entering the setup
menu and selecting one of the items of the “tESt” group (t.00
- t.08).
Note All the tests are made while in setup mode, where a time-out
is present. If no action is performed for about 5 minutes, the
mode is automatically exited and the instruments returns to
previous mode.
DISPLAY TEST
The display selftest procedure consists of lighting up all the
display segments together.
• To start the procedure select setup
item t.00 and an “OFF” blink-
ing message will appear on the
primary LCD.
• Press the (or ) key once and
the message will switch to a blink-
ing “GO”.
• Press CFM key to confirm or the
(or ) key again to return to
the previous status.
• If confirmation is given when the
“OFF” message is blinking, no ac-
tion is performed and it will move
to the next setup item (t.01).