• If an offset calibration has been
made, the instrument will turn to
“slope” calibration (as indicated by
the “SLO” message on the second-
ary display. The slope value is shown
on the primary LCD and the first digit
is blinking to permit modifications).
• Press the or key to modify the value or key to move
to the next digit.
• Once selected the desired value, press CFM to confirm.
• After confirmation the instrument will turn back to normal
mode.
• If the slope is invalid, the ”WRONG” indicator will blink
on the LCD.
Note Press LCD or CAL DATA key to exit calibration. The slope will
be set to the default value (57.5 mV/pH).
Note After direct selection of offset and slope, calibration data can
be viewed by pressing the CAL DATA key while in normal
mode, but no pH-buffer values will be displayed during the
data scrolling on the LCD.