OFFSET AND SLOPE DIRECT SELECTION
It is possible to edit directly the values of the offset and the
slope to calibrate the instrument.
• Press the CAL DATA key entering the last calibration data
scrolling and then press the SETUP key.
• A password entry is required. If a wrong password is con-
firmed, the instrument reverts back to the normal mode.
• Otherwise, if correct password is con-
firmed, the LCD will appears as
follows: the secondary LCD shows
“OFF” to indicate “offset” while the
primary one shows offset default
value. The first digit on the primary
LCD is blinking, and it is possible to
change it by pressing or key.
• Pressing the key will move to the
second digit while the first one will
be fixed.
• Pressing repeatedly the key will move cyclically through
the digits of the primary LCD.
• Once selected the desired value, press CFM to confirm the
calibration offset.
• If the offset is invalid, the “WRONG”
indicator will blink on the display.
Note If LCD or CAL DATA key is pressed before CFM, calibration
is aborted without changing the previous calibration data.