Calibration and Setup procedures are password protected.
Calibration: pH calibration at 2 points with two possible sets
of buffers (either 7.01 - 4.01 - 10.01 or 6.86 - 4.01 - 9.18),
or at 1 point with user-selectable value.
Temperature compensation of the HANNA standard buffers.
GLP features: old probe check, dead probe check, calibra-
tion time-out, last calibration data internally recorded (non-
volatile EEPROM memory): calibration date and time, pH
offset, pH slope, number of calibration points and corre-
spondent pH values.
Solution compensation of the pH reading.
Manual temperature setting when the temperature probe is
not inserted or temperature is out of range.
Input: pH/ORP electrode with BNC connector.
• Input: pH/ORP/Temperature Digital Transmitter (HI
504910) which can be connected up to 1.2 Km (0.5 miles)
far away.
Output (0-20 or 4-20 mA): one analog output for record-
ing of pH/ORP or controlling (only for models with PID).
RS485 communication with Short Messaging Service (SMS)
for GSM phones.
Data logging: 6000 samples pH/°C or ORP downloadable
through RS485 and HI92500 application software.
Real time clock.
OPTIONAL FEATURES (depending on model)
LEDs: four yellow LEDs are provided for signaling the ener-
gizing of relay 1, relay 2 (optional), relay 3 (optional) and
relay 4 (optional).
Relays: 1, 2 or 4 output relays. The relays LEDs are switched
ON whenever the correspondent relay is energized (NO
and COM connected).
Advanced Cleaning is provided through two cleaning commands
and variety of trigger events (relays 3 and 4 only).
Output (0-20 or 4-20 mA): second independent analog
output for recording of temperature or controlling (only for
models with PID).
PID control.