User Guide

56
All the alarm signals (red LED, alarm relay, fault currents) are
suspended while in hold mode (the correspondent error events
are not closed), unless the hold mode is being triggered by
one or more errors and no other trigger source (different from
an error) is active.
The analog output follows these rules:
if it is configured for control (i.e. setup item O.10 or O.20
set to “SEt”), its value is set to the minimum (e.g. 4 mA for
4-20 mA output);
if it is configured for a recorder, its value is either set to a
user-selected value through the setup item O.15 or O.25
(when item O.14 or O.24 is set to “USEr”) or frozen to the
last value, i.e. the output value just before entering the
hold mode (when item O.14 or O.24 is set to “HOLd”).
After the cause which made the instrument enter the hold
mode expires, the device exits hold mode but control and
alarms remain disabled for a delay which can be user-se-
lected (0 to 99s) through setup item C.70. In that situation
measurements are normally acquired, displayed and recorded
through the analog output or the RS485.
Note Alarms (red LED, alarm relay, fault currents) are not disabled
if the hold mode is being triggered by an error and no other
trigger source is active.