User Guide
55
HOLD MODE
This function is started by:
• calibration;
• setup;
• cleaning in place;
• the hold digital insulated input
1
when it is on; normally,
the signal level is polled at least every 4 seconds;
• the proper key combination ( and keys together) for
service; the same key combination is used both to start
and to stop the hold mode (the key combination acts as
the hold digital input). To activate the hold mode in this
way, password is required;
• the daily programmable control timing (see setup items
C.41 through C.57);
• an error event (see also the “Alarm - Error Configuration”
section);
• the hold start/stop RS485 command;
• the service hold (Calibration and Setup) in the Digital Trans-
mitter.
During hold mode, control and control relays are disabled.
If the meter is in idle or control mode
and displaying measurements, the last
measured value (both for temperature
and pH/ORP) is frozen on the display;
the secondary LCD shows “HLd”;
the temperature value can be displayed on the secondary
LCD by pressing and holding the key; when the key is
released the “HLd” message comes back on the secondary
LCD.
If the pH or mV or temperature values
are not available because the meter
did note take any measurement be-
fore going to hold mode, then the
primary or the secondary LCD shows
dashes.
1
There are two digital insulated inputs: one for hold mode and one for advanced cleaning.