IDLE MODE
During idle mode the device performs only measurements
but it does not activate relays in order to control the process
or let out a control signal to the analog output.
In a normal situation the alarm relay is energized (no alarm
condition) and the green LED is ON, the red LED is also fixed
ON to warn users the device is not controlling the process,
the yellow LEDs are OFF.
Moreover, the alarm relay could be de-energized upon an
error (whether that happens or not, depends on the custom-
ized alarm configuration; see “Alarm” section for more details).
Nevertheless, the error due to alarm threshold overtaking
does never generate an alarm during idle mode since all the
control functions have been disabled by the user.
The red LED flashes in any case upon an error.
Idle mode is useful to disable control actions when the exter-
nal devices are not properly settled or whenever any fault
circumstance is detected.
To set the meter to idle mode it is sufficient to disable the
control mode (setup item C.00 to “OFF”).
When measurements are in overflow, the upper or lower range
limit is shown on the LCD with blinking digits.