User Guide
53
Note If a cleaning session is being performed, it is possible to stop
it by pressing and holding the and keys together (
key first) or through RS485 by sending the appropriate com-
mand. When the advanced cleaning is aborted, in any case
a complete rinsing phase is performed before the actual ter-
mination of the cleaning action. If the abortion is issued dur-
ing the rinsing phase, the phase continues normally till its
natural end.
Note Calibration of whatever input or output can not be initiated
while the Simple or Advanced Cleaning is in progress. Con-
versely, cleaning can not be triggered while whatever cali-
bration is being performed.