User Guide

35
Code Valid Values Default Present
for ORP
t.03 Relays and LEDs test “OFF”: To skip without testing “OFF” yes
“GO”: To start the test
t.04 Analog output 1 test “OFF”: To skip without testing “OFF” yes
“GO”: To start the test
t.05 Analog output 2 test “OFF”: To skip without testing “OFF” yes
“GO”: To start the test
t.06 Hold digital input test “OFF”: To skip without testing “OFF” yes
“GO”: To start the test
t.07 Advanced cleaning “OFF”: To skip without testing “OFF” yes
digital input test “GO”: To start the test
Warning: cleaning actions could be triggered by the input going high during the
test (see note 23).
t.08 Watchdog test “OFF”: To skip without testing “OFF” yes
“GO”: To start the watchdog test
Notes (1): M1 can not be set to “OOHI” or “OOLO” if O.10 is set to
“SEt” and vice versa;
if M1 = “OOHI” then 16.00 pH or 2000 mV S S1+A1;
if M1 = “OOLO” then -2.00 pH or -2000 mV R S1-A1;
if M1 = “PIdH” then D1 R A1 and 16.00 pH or 2000 mVSS1+A1;
if M1 = “PIdL” then D1 R A1 and -2.00 pH or -2000 mV R S1-A1;
M2 can not be set to “OOHI” or “OOLO” if O.20 is set to “SEt”
and vice versa;
if M2 = “OOHI” then 16.00 pH or 2000 mV S S2+A2;
if M2 = “OOLO” then -2.00 pH or -2000 mV R S2-A2;
if M2 = “PIdH” then D2 R A2 and 16.00 pH or 2000 mVSS2+A2;
if M2 = “PIdL” then D2 R A2 and -2.00 pH or -2000 mV R S2-A2;
if M1 = “OOHI” and M2 = “OOLO” then S1-H1S S2+H2;
if M1 = “OOLO” and M2 = “OOHI” then S2-H2S S1+H1;
if M1 = “PIdH” and M2 = “OOLO” then S1S S2+H2;
if M1 = “OOLO” and M2 = “PIdH” then S1+H1R S2;