User Guide

34
Code Valid Values Default Present
for ORP
E.02 Max relay ON time 0 to 5 3 yes
error configuration and 24 to 29 (see note 7)
E.03 Life check error configuration 0 to 11 9 yes
(see note 11) and 24 to 35 (see note 7)
E.10 pH electrode broken 0 to 11 9 no
or leakage and 24 to 35 (see note 7)
E.11 Reference electrode broken 0 to 47 (see note 7) 21 yes
or dirty or not immersed
E.12 Old pH probe 0 to 11 0 no
error configuration and 24 to 35 (see note 7)
E.13 Dead pH probe 0 to 11 2 no
error configuration and 24 to 35 (see note 7)
E.14 Calibration time-out 0 to 11 0 no
error configuration and 24 to 35 (see note 7)
E.20 Temperature broken probe 0 to 11 3 no
error configuration and 24 to 35 (see note 7)
E.40 Digital Transmitter 0 to 11 9 yes
error configuration and 24 to 35 (see note 7)
E.50 Cellular error configuration 0 to 11 (see note 7) 3 yes
E.90 Power reset 0 to 11 2 yes
error configuration and 24 to 35 (see note 7)
E.91 EEPROM corruption 6 to 11 9 yes
error configuration and 30 to 35 (see note 7)
E.92 Watchdog error configuration 0 to 11 2 yes
and 24 to 35 (see note 7)
E.99 Level or pulse alarm relay “LE”: Level “LE” yes
signal (see note 13) “PULS”: Pulse
TEST (“tESt)
t.00 Display test “OFF”: To skip without testing “OFF” yes
“GO”: To start the display test
t.01 Keyboard test “OFF”: To skip without testing “OFF” yes
“GO”: To start the keyboard test
t.02 EEPROM test “OFF”: To skip without testing “OFF” yes
“GO”: To start the EEPROM test