User Guide

33
Code Valid Values Default Present
for ORP
SOLUTION COMPENSATION (“SOLC”)
S.00 Solution compensation enable “On”: compensation enabled “OFF” no
“OFF”: compensation disabled
S.10 pH value for point #1 -2.00 to 16.00 pH 7.00 pH no
S.11 Temperature value for -30 to 130.0 ºC 20 ºC no
point #1 (SCT1) |SCT1-SCT2| S 1.0 ºC
S.20 pH value for point #2 -2.00 to 16.00 pH 7.00 pH no
S.21 Temperature value for -30 to 130.0 ºC 30 ºC no
point #2 (SCT2) |SCT1-SCT2| S1.0 ºC
IN-LINE CLEANING (“CLEA”)
SIMPLE CLEANING (“SCLE”)
L.00 Rinsing time 5 to 99 s 20 s yes
L.01 Pause time 10 to 9999 minutes 1440 minutes yes
ADVANCED CLEANING (“ACLE”)
L.10 Pre-rinsing time 0 to 99 s 20 s yes
L.11 Detergent washing 0 to 99 s 10 s yes
L.12 Rinsing time 5 to 99 s 20 s yes
L.13 Pause time Min. pause time (L.14) to 1440 minutes yes
9999 minutes
L.14 Minimum pause time 10 to Pause time (L.13) 10 minutes yes
L.15 Cleaning trigger “ ti”: Timer only “ ti” yes
“ E”: Digital input or RS485 (external) only
“ti E”: Timer and digital input or RS485 (external)
“tiEM”: Timer masked by digital input (external)
L.16 Repeated cleaning number 0 to 10 0 yes
L.17 Number of cleaning 0 to 10 0 yes
without detergent
ERROR CONFIGURATION (“Erro”)
E.00 Alarm for setpoint 1 0 to 5 3 yes
error configuration and 24 to 29 (see note 7)
E.01 Alarm for setpoint 2 0 to 5 5 yes
error configuration and 24 to 29 (see note 7)