User Guide
30
Code Valid Values Default Present
for ORP
0.05 Hold digital output “OFF”: Disabled “HOLd” yes
“HOLd”: Enabled upon hold mode
ANALOG OUTPUT #1 (“Out1”)
O.10 Analog output 1 mode “rECO”: Recorder “rECO” yes
(see notes 6, 21) “SEt”: Control, setpoint 1
O.11 Analog output 1 type “0-20”: 0-20 mA 4-20 (6) yes
“4-20”: 4-20 mA
O.12 Measurement value for -2.00 to 16.00 pH 0.00 pH or yes
analog output 1 minimum or -2000 to 2000 mV -2000 mV (6)
(O_VARMIN1) (O_VARMIN1 R O_VARMAX1-1 pH or 50 mV,
O_VARMIN1 R O_HOLD1R O_VARMAX1)
O.13 Measurement value for -2.00 to 16.00 pH 14.00 pH or yes
analog output 1 maximum or -2000 to 2000 mV 2000 mV (6)
(O_VARMAX1) (O_VARMIN1 R O_VARMAX1-1 pH or 50 mV,
O_VARMIN1 R O_HOLD1R O_VARMAX1)
O.14 Analog output 1 value upon “USEr”: User selected value “HOLd” yes
hold mode (see note 16) “HOLd”: Previous value is frozen
O.15 Analog output 1 value upon -2.00 to 16.00 pH 7.00 pH or yes
hold mode if O.14=”USEr” or -2000 to 2000 mV 0 mV (6)
(O_HOLD1) (it must be within the O_VARMIN1
to O_VARMAX1-1 interval)
ANALOG OUTPUT #2 (“Out2”)
O.20 Analog output 2 mode “rECO”: Recorder (temperature) “rECO” (pH) yes
(see notes 21, 26) “SEt”: Control, setpoint 2 (pH) or “OFF” (ORP)
or
“OFF”: Disabled
“SEt”: Control, setpoint 2 (ORP)
O.21 Analog output 2 type “0-20”: 0-20 mA “4-20” yes
“4-20”: 4-20 mA
O.22 Measurement value for -30 to 130.0 ºC 0 ºC no
analog output 2 minimum (O_VARMIN2 R O_VARMAX2-10 ºC,
(O_VARMIN2) O_VARMIN2 R O_HOLD2R O_VARMAX2)
O.23 Measurement value for -30 to 130.0 ºC +100.0 ºC no
analog output 2 maximum (O_VARMIN2 R O_VARMAX2-10 ºC,
(O_VARMAX2) O_VARMIN2 R O_HOLD2R O_VARMAX2)