User Guide

22
is exited (by pressing the CAL key).
• When performing the second calibration point, it is pos-
sible to choose between two values, 25°C and 50°C. Pressing
the or key the value on the secondary LCD will switch
between the two possibilities.
• Once selected the value, immerse the temperature probe
in the second beaker as near as possible to the Checktemp
and the calibration procedure will be the same as for the
first point.
VOLT CALIBRATION
The instrument is factory calibrated; however the user may
also perform the Volt calibration, which is a procedure avail-
able for the meter configured as pH controller only.
During pH measurements the instrument reads a mV value
from the electrode and then converts it into a pH value.
• Set item I.04 to “OFF” to disable the Matching Pin.
• Connect a HI 931001 or HI 8427 simulator to the BNC
socket.
• Once entered the calibration mode,
move through the menu with or
key, select the Volt calibration (the
primary LCD shows “UOLt” mes-
sage) and confirm it by pressing the
CFM key.
• After confirmation the primary LCD
will show the actual mV value and
the secondary LCD will display the
first calibration point.
The calibration is performed over two points, 0 and 350 mV.
• Set the HI 931001 or HI 8427 simulator to 0 mV.
• When the reading becomes stable, if the measured value
is close to the calibration point the CFM tag starts blink-
ing, otherwise the WRONG indicator will flash.
• In the first case press the CFM key to confirm calibration.