User Guide

20
TEMPERATURE CALIBRATION
• Prepare a beaker containing
crushed ice and water at 0°C (32°F)
and another one with hot water at
25°C (77°F) or 50°C (122°F).
• Use a Checktemp or another cali-
brated thermometer with a
resolution of 0.1° as a reference
thermometer.
• Immerse the temperature probe in
the beaker with ice and water as
near to the Checktemp as possible.
0 °C
(32 °F)
°C
50 ºC
(122 ºF)
°C
0 °C
(32 °F)
°
C
• Once visualized the selected output, press the key to
choose the range of the analog output (0-20 mA or 4-20
mA); then press the CFM key to confirm the choice.
• Once selected and confirmed the
range for the analog output, the
secondary LCD shows the first point
of calibration (1 or 4 mA) and the
primary LCD displays the range of
the actual calibration.
• With the tester check the real current value provided by the
instrument at the output port. If this value is different from
the expected one, it is possible to change the current
erogated by the instrument pressing the or key.
• Adjust the value until it matches with the first point of cali-
bration, then press CFM to confirm. The instrument will
turn to the second calibration point (20 mA).
• The calibration procedure is the same as for the first point.
• Once the instrument is calibrated for both points (and con-
firmed), it will exit from the calibration menu.