User Guide

19
• When the reading becomes stable, if the ORP value is
close to the calibration point, the “CFM” indicator starts
blinking; otherwise the “WRONGindicator blinks and
the “CAL” is fixed on.
• In the first case press CFM to confirm calibration. The
meter will proceed showing the scrolling message “Press
CFM again to confirm the current buffer or right to es-
cape” (to prevent from confirming the calibration point
inadvertently). Pressing again CFM the secondary LCD will
display the second calibration point.
• In the second case (blinking WRONG”) the meter will
remain in the WRONG state until the reading becomes
unstable or the calibration mode is exited by pressing the
CAL key.
• For the second calibration point it is possible to choose
between 350 mV or 1900 mV. Pressing the or key the
value on the secondary LCD will switch between the two
possibilities.
• Once selected the second calibration point, set the HI
931001 (350 mV) or HI 8427 (350 or 1900 mV) simula-
tor to the same value and the calibration proceeds as for
the first point.
Note In ORP calibration there is no time-out.
ANALOG OUTPUT CALIBRATION
The instrument can be provided with one or two analog out-
puts, each of them can be calibrated.
Connect an Ammeter or the HI 931002 calibrator to the
analog output to measure the current erogated by the meter.
• Enter the calibration mode and move through the menu
using the or key until “Out 1” or “Out 2” message
appears on the primary LCD.