• The pH calibration value will switch
to 7.00, the first digit starts blinking
and it is possible to change its value
simply using the or key.
• Once selected the first digit value, press the key: the first
digit will be fixed and the second one will start blinking.
Pressing of the key repeatedly will result on circularly
moving on the secondary LCD.
• When the desired calibration value is reached (must be
within 0.00 to 16.00 pH), press the CFM key to confirm
and the calibration will proceed as described above.
• If the selected value is outside boundaries, the confirma-
tion is not accepted and the first digit keeps blinking (waiting
for confirmation of a valid value).
Note If SETUP key is pressed instead of CFM, the calibration value
selection is aborted and the meter reverts back to a two-point
calibration.
ORP CALIBRATION
To perform ORP calibration it is necessary to connect an HI
931001 or HI 8427 simulator to the BNC socket.
The meter has to be set as ORP controller.
• Set item I.04 to “OFF” to disable the Matching Pin.
• A two-point calibration has to be performed: the first point
value is 0 mV and the second one can be chosen between
350 mV and 1900 mV.
• Enter the calibration mode, select the ORP calibration (use
the and keys to move through setup menu) and press
the CFM key.
• Set the HI 931001 or HI 8427 simulator to 0 mV.
• The primary LCD will display the cur-
rent mV measure and the secondary
LCD will show the first calibration
point (0 mV).