• In the second case (pH value not close to the buffer) the
meter will remain in the same state until the reading be-
comes unstable or the calibration mode is quitted.
• For the second buffer value it is possible to choose be-
tween pH 4.01 and pH 10.01 (or pH 4.01 and pH 9.18 if
the NIST set has been selected). Use the or key to
switch between the two possibilities.
• Once selected the buffer, the procedure is the same as for
the first calibration point.
Note A time-out of 1.5 minutes is present for the pH electrode
response time. During calibration, if the pH reading is not
stable after 1.5 minutes, the device displays twice the scroll-
ing message “time-out”, then shows “WRONG” and it is not
possible to complete calibration.
• At the end of calibration, with the meter set as pH control-
ler, the instrument checks if the offset is between -30 and
30 mV and the slope between 53.5 and 62 mV/pH. If the
values are not within these ranges, the message “OLd
ProbE” scrolls twice across the LCD. The electrode is still
working, but it is necessary to perform a cleaning proce-
dure (see “Electrode conditioning and maintenance” and
“In-line Cleaning” sections) or replace it.
If the offset is outside the -60 to 60 mV range, the “dEAd
ProbE” message will scroll across the LCD and the corre-
sponding error is activated. The electrode has to be replaced
as soon as possible because there is no reliability on the
measured pH values.
One-point calibration
A one-point calibration at a value different from the standard
buffer values is possible by entering the desired calibration
value. This is the actual pH value at the current calibration
temperature.
• Enter the pH calibration mode (no matter if the standard or
NIST buffer set is selected), then press the
SETUP key while the secondary LCD is dis-
playing the first buffer value (pH 7.01 or pH
6.86).