User Guide

12
INSTALLATION
Input power: Connect a 3-wire power cable
to the terminal strip line (L), earth (PE) and
neutral (N) terminal connections.
Power: 100VAC-120mA / 115VAC-100mA / 230VAC-50mA.
Line Contact: 400mA fuse inside.
PE must be connected to ground; leakage current 1mA.
Refer to diagram on page #9