Code Valid Values Default Present
for ORP
TEST (“tESt”)
t.00 Display test “OFF”: To skip without testing “OFF” yes
“GO”: To start the display test
t.01 Keyboard test “OFF”: To skip without testing “OFF” yes
“GO”: To start the keyboard test
t.02 EEPROM test “OFF”: To skip without testing “OFF” yes
“GO”: To start the EEPROM test
t.03 Relays and LEDs test “OFF”: To skip without testing “OFF” yes
“GO”: To start the test
t.04 Analog output 1 test “OFF”: To skip without testing “OFF” yes
“GO”: To start the test
t.05 Analog output 2 test “OFF”: To skip without testing “OFF” yes
“GO”: To start the test
t.06 Hold digital input test “OFF”: To skip without testing “OFF” yes
“GO”: To start the test
t.07 Advanced cleaning “OFF”: To skip without testing “OFF” yes
digital input test “GO”: To start the test
Warning: cleaning actions could be triggered by the input going high during the
test (see note 23).
t.08 Watchdog test “OFF”: To skip without testing “OFF” yes
“GO”: To start the watchdog test
Notes (1): M1 can not be set to “OOHI” or “OOLO” if O.10 is set to
“SEt” and vice versa;
if M1 = “OOHI” then 16.00 pH or 2000 mV
S S1+A1;
if M1 = “OOLO” then -2.00 pH or -2000 mV
R S1-A1;
if M1 = “PIdH” then D1
R A1 and 16.00 pH or 2000 mVS S1+A1;
if M1 = “PIdL” then D1
R A1 and -2.00 pH or -2000 mV R S1-A1;
M2 can not be set to “OOHI” or “OOLO” if O.20 is set to “SEt”
and vice versa;
if M2 = “OOHI” then 16.00 pH or 2000 mV
S S2+A2;
if M2 = “OOLO” then -2.00 pH or -2000 mV
R S2-A2;
if M2 = “PIdH” then D2
R A2 and 16.00 pH or 2000 mVS S2+A2;
Code Valid Values Default Present
for ORP
ERROR CONFIGURATION (“Erro”)
E.00 Alarm for setpoint 1 0 to 5 3 yes
error configuration and 24 to 29 (see note 7)
E.01 Alarm for setpoint 2 0 to 5 5 yes
error configuration and 24 to 29 (see note 7)
E.02 Max relay ON time 0 to 5 3 yes
error configuration and 24 to 29 (see note 7)
E.03 Life check error configuration 0 to 11 9 yes
(see note 11) and 24 to 35 (see note 7)
E.10 pH electrode broken 0 to 11 9 no
or leakage and 24 to 35 (see note 7)
E.11 Reference electrode broken 0 to 47 (see note 7) 21 yes
or dirty or not immersed
E.12 Old pH probe 0 to 11 0 no
error configuration and 24 to 35 (see note 7)
E.13 Dead pH probe 0 to 11 2 no
error configuration and 24 to 35 (see note 7)
E.14 Calibration time-out 0 to 11 0 no
error configuration and 24 to 35 (see note 7)
E.20 Temperature broken probe 0 to 11 3 no
error configuration and 24 to 35 (see note 7)
E.21 Temperature level 0 to 5 3 no
error configuration and 24 to 29 (see note 7)
E.40 Digital Transmitter 0 to 5 9 yes
error configuration and 24 to 35 (see note 7)
E.50 Cellular error configuration 0 to 11 (see note 7) 3 yes
E.90 Power reset 0 to 11 2 yes
error configuration and 24 to 35 (see note 7)
E.91 EEPROM corruption 6 to 11 9 yes
error configuration and 30 to 35 (see note 7)
E.92 Watchdog error configuration 0 to 11 2 yes
and 24 to 35 (see note 7)
E.99 Level or pulse alarm relay “LE”: Level “LE” yes
signal (see note 13) “PULS”: Pulse