Code Valid Values Default Present
for ORP
SOLUTION COMPENSATION (“SOLC”)
S.00 Solution compensation enable “On”: compensation enabled “OFF” no
“OFF”: compensation disabled
S.10 pH value for point #1 -2.00 to 16.00 pH 7.00 pH no
S.11 Temperature value for -30 to 130.0 ºC 20 ºC no
point #1 (SCT1) |SCT1-SCT2| S 1.0 ºC
S.20 pH value for point #2 -2.00 to 16.00 pH 7.00 pH no
S.21 Temperature value for -30 to 130.0 ºC 30 ºC no
point #2 (SCT2) |SCT1-SCT2| S 1.0 ºC
IN-LINE CLEANING (“CLEA”)
SIMPLE CLEANING (“SCLE”)
L.00 Rinsing time 5 to 99 s 20 s yes
L.01 Pause time 10 to 9999 minutes 1440 minutes yes
ADVANCED CLEANING (“ACLE”)
L.10 Pre-rinsing time 0 to 99 s 20 s yes
L.11 Detergent washing 0 to 99 s 10 s yes
L.12 Rinsing time 5 to 99 s 20 s yes
L.13 Pause time Min. pause time (L.14) to 1440 minutes yes
9999 minutes
L.14 Minimum pause time 10 to Pause time (L.13) 10 minutes yes
L.15 Cleaning trigger “ ti”: Timer only ti” yes
“ E”: Digital input or RS485 (external) only
“ti E”: Timer and digital input or RS485 (external)
“tiEM”: Timer masked by digital input (external)
L.16 Repeated cleaning number 0 to 10 0 yes
L.17 Number of cleaning 0 to 10 0 yes
without detergent
TEMPERATURE LEVEL ALARM (“tEMP”)
U.00 Max. temperature level -30 to 130 °C 130.0 °C no
(see note 32)
U.01 Min. temperature level -30 to 130 °C -30.0 °C no
(see note 32)
Code Valid Values Default Present
for ORP
r.02 Current year 2000 to 2099 (see note 20) from RTC (5) yes
r.03 Current time 00:00 to 23:59 (see note 20) from RTC (5) yes
CELLULAR/MODEM/PC CONNECTION (“PHOn”)
P.00 RS485 Connection type “PC”: PC or modem connection “PC” yes
(see note 20) “CELL”: Cellular module connection
P.01 PIN Number (see note 20) 0000 to 9999 0000 yes
P.02 Telephone number #1 00000000000000000000 - - - - - - - - - - - - - - - - - - - -
(see notes 20, 27) to 99999999999999999999 yes
(“- - - - - - - - - - - - - - - - - - - -” indicates no number)
P.03 Telephone number #2 00000000000000000000 - - - - - - - - - - - - - - - - - - - -
(see notes 20, 27) to 99999999999999999999 yes
(“- - - - - - - - - - - - - - - - - - - -” indicates no number)
P.04 Number of remaining 000 to 200 100 yes
messages (see notes 20, 28) and 222
P.05 Number of repeated 0 to 5 2 yes
SMSs (see notes 20, 29)
P.06 Delay between two 05 to 60 minutes 10 yes
subsequent SMSs (see notes 20, 30)
P.07 SIM expiration day 01 to 31 01 yes
(see notes 20, 31)
P.08 SIM expiration month 01 to 12 01 yes
(see notes 20, 31)
P.09 SIM expiration year 2000 to 2099 2010 yes
(see notes 20, 31)
READING OFFSETS (“OFFS”)
F.00 pH or ORP actual value measured value -1.00 pH or measured yes
(see notes 17, 25) -200 mV to measured value value
+1.00 pH or +200 mV
F.01 pH or ORP reading offset -1.00 to +1.00 pH or 0.00 pH yes
adjustment (see note 25) -200 to +200 mV or 0 mV
F.10 Temperature actual value measured value -10.0 ºC to meas. value no
(for ATC only, see notes 18, 25) measured value +10.0 ºC (see note 17)
F.11 Temperature reading offset -10.0 to +10.0 ºC 0.0 ºC no
adjustment (for ATC only, see notes 18, 25)