Instruction Manual

53
If the standard solution concentration is validated,
will appear on the LCD. Press to update
calibration. The calibration point value will be added to the Calibrated Standards section.
The “Please wait...” message will appear on the LCD until the reading is stable and 10 seconds elapsed,
time in which immerse the Ion Selective Electrode and the temperature probe into the next standard solution
and follow the above procedure or press to exit calibration.
Notes: The new added calibration point will replace an old
one if the difference between them is less than 20%
of the standard solution.
If the existing stored calibration is full (five calibra-
tion points), a pop-up menu will be displayed on the
LCD in which you can select with or
the standard solution you want to replace with the
current one. Press to delete the selected
calibrated point and then press to update
calibration with the new standard solution.
If the isopotential point of the electrode is unknown, the
ion calibration and measurements can be performed only
without temperature compensation (see ISE Setup, Tem-
perature option for details, page 33). When the temperature
compensation option is disabled, at least a two-point
calibration must be performed in order to perform ISE
measurements. Otherwise, “-----” will appear on the LCD.
When in MTC mode, if pressing after enter-
ing ISE calibration and while a standard solution is
selected, a pop-up menu will be displayed on the LCD
in which the concentration and the temperature value
can be ajusted by pressing and then
or keys. Press to save the modi-
fied value and then / to select
next /previous value to be adjusted.
CALIBRATION MESSAGES
Wrong standard solution. Please check the standard solution: this message appears when the
difference between the ppm reading and the value of the selected standard solution concentration (ppm)
64
ConductivityConductivity
ConductivityConductivity
Conductivity
SETUP SETUP
SETUP SETUP
SETUP
The Conductivity Setup menu allows the user to set the parameters related to the conductivity measurement.
Accessing Conductivity Setup
Press while in
Measure
mode and then to
select the
Conductivity
measure mode.
Press and then to access
Conductivity
Setup
menu.
To access a conductivity setup options:
Use or to highlight the desired option.
Press to access the selected option or to
exit setup.
The following is a detailed description of the
Conductivity Setup
option screens.
Profile
Choosing this option the measuring and the calibration mode can be customized. Up to 10 profiles can be
defined by the user.
The available options are:
Save Current Profile:
save the current profile.
Load Profile:
load from available profiles.
Delete Profile:
delete a profile.