Instruction Manual
50
Relative mVRelative mV
Relative mVRelative mV
Relative mV
MEASUREMENTS MEASUREMENTS
MEASUREMENTS MEASUREMENTS
MEASUREMENTS
To measure the Relative mV of a sample:
• Press
and then to enter
Relative mV
Measure
mode.
• The instrument will display the measured Relative mV value
on the LCD, together with a short GLP information about the
last calibration or “Not Calibrated” message if no Rel mV
calibration was performed (no Rel mV offset set).
Notes: • If the measured mV potential is out of range, “-----”
will be displayed on the LCD.
• If pressing while in
Relative mV Measure
mode, the Relative mV offset screen will be displayed
on the LCD. Press to clear the existing
Relative mV or use or to adjust
the Relative mV value and then press
to save
the current Relative mV. Press at any time
to return to
Relative mV Measure
mode.
• Before performing a Rel mV calibration, make sure
that the electrode is immersed into the sample to be
measured.
67
Temperature
From the
Temperature
menu the user can choose the temperature
source and units, as well as the temperature compensation mode,
reference temperature and compensation coefficient.
To access a
Temperature
option:
• Use or to highlight the
Temperature
op-
tion from the
Conductivity Setup
menu.
• Press
to access the
Temperature
option.
Temperature Source
To set the
temperature source:
• Use or to highlight the
Temperature Source
option.
• Press
and then use or to select
Automatic
or
Manual
temperature source.
• Press
to confirm your selection or press to
cancel operation.