Instruction Manual

49
mV/ORPmV/ORP
mV/ORPmV/ORP
mV/ORP
MEASUREMENTS MEASUREMENTS
MEASUREMENTS MEASUREMENTS
MEASUREMENTS
Oxidation-reduction potential (ORP) measurements provide the quantification of the oxidizing or reducing power of the
tested sample.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.
DIRECT MEASUREMENT
To measure the mV of a sample using the Direct reading mode:
Press and then to enter
mV Measure
mode.
Select the Direct reading mode (see
mV Setup
for details).
Submerge the tip of the ORP electrode (4 cm/1½”) into the
sample to be tested and allow a few seconds for the reading
to stabilize.
The instrument will display the measured mV value on the LCD.
Note: If the reading is out of range, “-----” will be displayed
on the LCD.
DIRECT/AUTOHOLD MEASUREMENT
To measure mV of a sample using the Direct/AutoHold reading mode:
Press and then to select
mV Measure
mode.
Select the Direct/AutoHold reading mode (see
mV Setup
for
details).
Submerge the tip of the ORP electrode approximately 4 cm
(1½”) into the sample to be tested and allow a few
seconds for the reading to stabilize.
The measured mV value will be displayed on the LCD. If
pressing , the “AutoHold” indicator will start blinking
on the LCD until the stability criterion is reached. The mV value
will be frozen on the LCD, along with “AutoHold” indicator.
To return to normal
Measure
mode press .
Note: If the reading is out of range, “-----” will be displayed
on the LCD.
mV & Relative mV MEASUREmV & Relative mV MEASURE
mV & Relative mV MEASUREmV & Relative mV MEASURE
mV & Relative mV MEASURE
MENTMENT
MENTMENT
MENT
SS
SS
S
68
Temperature Compensation
The user can choose from the following options:
Linear -
the meter will compensate automatically the conductivity using the following formula:
where:
C
ref
- conductivity at reference temperature
- compensation coefficient
T - temperature
ref - reference temperature
Non-Linear -
recommended for measuring the conductivity of the natural water in accordance with the compensation
table on the next page.
Disabled -
the meter will display the conductivity with no temperature compensation.
To set the
temperature compensation
mode:
Use or to highlight the
Temperature Com-
pensation
option.
Press and then use or to select
Linear, Non-Linear
or
Disabled
option.
Press to confirm your selection or press to cancel
operation.
Whatever form of compensation is used, the reading will not be as accurate as taking a reading of the sample’s
conductivity at the reference temperature.