User Manual

70
Notes: “Err3” is sent if the Log on demand is empty.
“Err4” is sent if the requested parameter is not available.
“Err6” is sent if the requested range is not available.
“Err8” is sent if the instrument is not in measurement mode.
Invalid commands will be ignored.
The reading status (1 char): R, O, U
The unit of the reading (2 chars): it is sent only for
ISE, EC, Resistivity, TDS, Salinity log
The calculated reading, with sign, decimal point
and exponent (12 chars)
The temperature reading, with sign and decimal
point (7 chars)
The reading status (1 char): R, O, U; not available
for EC log
The second unit of the second reading (2 chars); it is
sent only for Resistivity, TDS and Salinity log
The second reading, with sign and decimal point
(8 chars); not available for EC log
The logged time, yymmddhhmmss (12 chars)
The offset, with sign and decimal point (7 chars)
(not available for ISE log)
The slope for Channel 1 and the cell constant for
Channel 2, with sign and decimal point (7 chars)
(not available for Rel mV log)
The temperature probe presence (1 char)
For channel 2 only
The TDS factor (for the TDS log only) or the salinity
factor (for the salinity log only), with sign and decimal
point (6 chars)
The temperature reference value (2 chars)
The temperature compensation mode (2 chars)
The temperature compensation coefficient, with sign
and decimal point (6 chars)