User Manual

69
For each record:
The number of the record (3 chars)
The reading status (1 char): R, O, U
The unit of the reading (2 chars): it is sent only
for ISE, EC, Resistivity, TDS, Salinity log only
The calculated reading, with sign, decimal point
and exponent (12 chars)
The temperature reading, with sign and decimal
point (7 chars)
The second reading status (1 char): R, O, U; not
available for EC log
The second unit of the reading (2 chars); it is sent
only for Resistivity, TDS and Salinity log only
The second reading, with sign and decimal point
(8 chars); not available for EC log
The logging time (6 chars): it is sent only for
AutoEnd log
The temperature probe persence (1 char)
LODxPyyy Requests the yyy
th
pH record logged data when x=1
LODxMyyy Requests the yyy
th
mV/Rel mV record logged data when x=1
LODxIyyy Requests the yyy
th
ISE record logged data when x=1
LODxEyyy Requests the yyy
th
EC record logged data when x=2
LODxRyyy Requests the yyy
th
Resistivity record logged data when x=2
LODxTyyy Requests the yyy
th
TDS record logged data when x=2
LODxNyyy Requests the yyy
th
Salinity record logged data when x=2
LODxPALL Requests all pH Log on demand when x=1
LODxMALL Requests all mV/Rel mV Log on demand when x=1
LODxIALL Requests all ISE Log on demand when x=1
LODxEALL Requests all EC Log on demand when x=2
LODxRALL Requests all Resistivity Log on demand when x=2
LODxTALL Requests all TDS Log on demand when x=2
LODxNALL Requests all Salinity Log on demand when x=2
The answer string contains:
The number of records (3 chars): this is sent only is
“ALL” data is requested
The log type (2 chars)
The number of the record (3 chars)