User Manual

67
encoding
0x01 - beep ON (else OFF)
0x02 - degrees Celsius (else Fahrenheit)
0x04 and 0x08 - this 2 bits represent the
temperature compensation mode (10 - ATC,
01 - MTC, 00 - NoTC)
The cell constant, with sign and decimal point (7 chars)
The temperature compensation coefficient, with sign
and decimal point (6 chars)
The temperature reference value (2 chars)
The TDS factor, with sign and decimal point (5 chars)
The short name of the selected language (3 chars)
NSLxy Requests the number of logged samples (4 chars).
The command parameter (2 chars):
x=1 request for channel 1
x=2 request for channel 2
y=P request for pH range
y=M request for mV and Rel mV ranges
y=I request for ISE range
y=E request for EC range
y=R request for Resistivity range
y=T request for TDS ranges
y=N request for NaCl range
LLSxy Requests the number of lots (information about lots) on the specified
channel and range (x - channelnumber; y - range first letter)
x=1 request for channel 1
x=2 request for channel 2
y=P request for pH range
y=M request for mV and Rel mV ranges
y=I request for ISE range
y=E request for EC range
y=R request for Resistivity range
y=T request for TDS ranges
y=N request for NaCl range
The answer string contains:
number of lots (3 chars)
For each lot:
the lot number (3 chars)
the number of records in lot (3 chars)