User Manual

66
sign and decimal point (6 chars)
Salinity calibration data (if available), which contains:
the number of calibrated standards (1 char always 1)
the salinity factor, with sign and decimal point
(7 chars)
calibration time, yymmddhhmmss (12 chars)
PARx Requests the setup parameters setting.
x=1 setup parameters for channel 1
x=2 setup parameters for channel 2
The answer string contains:
Lot log interval index (2 ASII chars - it represents
the index from the setup list for the log interval)
Backlight Value (1 ASCII char)
Contrast Value (2 ASCII char)
Instrument ID (4 chars)
Channel 1 data
Calibration Alarm time out for pH (2 chars)
Calibration Alarm timeout for ISE (2 chars)
SETUP information (2 chars): 8 bit hexadecimal
encoding.
0x01 - beep ON (else OFF)
0x02 - degrees Celsius (else degrees Fahrenheit)
0x04 - Offset calibration (else Point calibration)
The number of pH custom buffers (1 char)
The custom buffer values, with sign and decimal
point, for each defined custom buffer (7 chars)
The ID of the ISE electrode (2 chars) (standard or
custom)
The molar weight of the selected ION, with sign and
decimal point (9 ASCII characters)
The electrode slope (6 ASCII chars)
The ion charge (2 chars)
The ISE unit (2 chars)
The short name of the selected language (3 chars)
Channel 2 data
Calibration alarm timeout for the EC (2 chars)
Calibration alarm timeout for the Salinity (2 chars)
Setup information (2 chars): 8 bit hexadecimal