electrode condition, with sign (3 chars). The
“-01” code means not calculated
ISE calibration data (if available) which contains:
the number of calibrated standards (1 char)
the ISE unit (2 chars)
the calibration slope, with sign and decimal point
(7 chars)
the calibration time of the standard, yymmdd
hhmmss (12 chars)
standards information (for each standard)
type (1 char): 0 - always standard solution
status (1 char): N (new) - calibrated in last
calibration; O (old) - from an old calibration
warnings during calibration (2 chars): 00 -
no warning
standard value, with sign and decimal point
and exponent (12 chars)
calibration time, yymmddhhmmss (12 chars)
Rel mV calibration data (if available), which contains:
the calibration offset, with sign (7 chars)
the calibration time, yymmddhhmmss (12 chars)
Channel 2 GLP
EC calibration data (if available), which contains:
the number of calibrated standards (1 char)
the offset, with sign and decimal point (7 chars)
the cell constant, with sign and decimal point (7 chars)
the calibration time, yymmddhhmmss (12 chars)
standards information (for each buffer)
the unit of the standard (0 - microsiemens,
1 - milisiemens)
the standard value, with sign, decimal
point and exponont (11 chars)
the calibration time of the standard,
yymmddhhmmss (12 chars)
the temperature reference value (2 chars)
the temperature compensation mode (2 chars:
00 - NoTC, 01 - MTC, 02 - ATC)
the temperature compensation coefficient, with