User Manual

64
The unit of the second reading (1 char) (it is sent
only for Resistivity, TDS and Salinity ranges only). It
represents the unit of the EC reading (0 - microsiemens,
1 - milisiemens)
Secondary reading (only when primary reading is
not mV or EC) - 8 ASCII chars, including sign and
decimal point.
Temperature reading - 7 ASCII chars, with sign and
decimal point, always in ºC.
MDR Requests the instrument model name and firmware code
(16 ASCII chars).
GLPx Requests the calibration data record.
x=1 calibration data from channel 1
x=2 calibration data from channel 2
The answer string contains:
GLP status (1 char): represents a 4 bit hexadecimal
encoding.
0x01 - pH calibration available (Ch1) or EC
calibration available (Ch2)
0x02 - Rel mV calibration available (Ch1) or
Salinity calibration available (Ch2)
0x04 - ISE calibration available
Channel 1 GLP
pH calibration data (if available), which contains:
the number of calibrated buffers (1 char)
the offset, with sign and decimal point (7 chars)
the average of slopes, with sign and decimal
point (7 chars)
the calibration time, yymmddhhmmss (12 chars)
buffers information (for each buffer)
type (1 char): 0 - standard, 1 - custom
status (1 char): N (new) - calibrated in last
calibration; O (old) - from an old calibration.
warnings during calibration (2 chars): 00 -
no warning, 04 - Clean Electrode warning
buffer value, with sign and decimal point and
exponent (12 chars)
the calibration time of the buffer, yymmdd
hhmmss (12 chars).