User Manual

63
Resistivity, EC and temperature reading on the
Resistivity range
TDS, EC and temperature reading on the Resistivity
range
Salinity, EC and temperature reading on the
Resistivity range
The answer string contains:
Meter mode (2 chars):
10 pH range/0.001 resolution on channel 1
11 pH range/0.01 resolution on channel 1
12 pH range/0.1 resolution on channel 1
• 13 mV range on channel 1
• 14 Relative mV range on channel 1
• 15 ISE range on channel 1
• 20 EC range on channel 2
• 21 Resistivity range on channel 2
• 22 TDS range on channel 2
• 23 Salinity range on channel 2
Meter status (2 chars of status byte): represents a 8
bit hexadecimal encoding.
0x10 - temperature probe is connected
0x01 - new GLP data available
0x02 - new SETUP parameter
0x04 - out of calibration range
0x08 - the meter is in autoend point mode
Reading status (2 chars): R - in range, O - over
range, U - under range. First character corresponds
to the primary reading. Second character corresponds
to the second reading (if second reading exists).
The unit of the reading (2 chars) (for ISE, EC, Resis-
tivity, TDS, Salinity ranges only)
for ISE range it represent the index of the unit
selected from setup (0-ppt, 11-User)
for EC range: 0 - microsiemens, 1 - milisiemens
for TDS range: 0 - ppm, 1 - g/L
for Salinity range the unit represents always %
Primary reading (corresponding to the selected range)
- 12 ASCII chars, including sign and decimal point
and exponent.