CHRxx Change the instrument range according with the parameter
value (xx):
xx=10 pH range/0.001 resolution on channel 1
xx=11 pH range/0.01 resolution on channel 1
xx=12 pH range/0.1 resolution on channel 1
xx=13 mV range on channel 1
xx=14 Relative mV range on channel 1
xx=15 ISE range on channel 1
xx=20 EC range on channel 2
xx=21 Resistivity range on channel 2
xx=22 TDS range on channel 2
xx=23 Salinity range on channel 2
The instrument will answer for these commands with:
<STX> <answer
>
<ETX>
where: <STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized com-
mand)
<CAN> is 24 ASCII code character (corrupted command)
COMMANDS REQUIRING AN ANSWER
The instrument will answer for these commands with:
<STX> <answer> <checksum> <ETX>
where the checksum is the bytes sum of the answer string sent as 2 ASCII
characters.
All the answer messages are with ASCII characters.
RAS Causes the instrument to send a complete set of readings in
according with the current range:
pH, mV reading and temperature on pH range.
mV and temperature reading on mV range
Rel mV, absolute mV and temperature reading on
Rel mV range.
concentration, mV and temperature reading on ppm
range.
EC and temperature reading on the EC range