User Manual

59
Press CAL key on the pH range or on the EC range. The calibration
screen is displayed. Press Temp function key to enter the temperature
calibration mode. The temperature calibration is performed for the
current channel only. The calibration routine is the same for both
channels, so Ch1 calibration is presented next.
Prepare a vessel containing ice and water and another one containing
hot water (at approximately
50 ºC or 122 ºF). Place insulation
material around the vessels to minimize temperature changes.
Use a calibrated thermometer with a resolution of 0.1 ºC as a
reference thermometer. Connect the temperature probe or the EC
probe to the appropriate socket.
Immerse the temperature probe or the EC probe including temperature
sensor into the vessel with ice and water as close as possible to the
reference thermometer. Allow a few seconds for the probe to stabilize.
Use the ARROW keys to set the calibration point value to that of ice
and water mixture, measured by the reference thermometer. When
the reading is stable and close to the selected calibration point, the
Confirm function key is displayed.
Press Confirm to confirm.
The second expected calibrated point is displayed.
Immerse the temperature probe into the second vessel as close as
possible to the reference thermometer. Allow a few seconds for the
probe to stabilize.
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
(for technical personnel only)