User Guide

6
Rear Panel
1. Socket for D.O. Probe
2. Reset Button
3. RS 232C Connector
4. DC Power Socket (for HI 710005 or HI 710006)
35
19) etx end
?DM Requests the meter to send the selected lot data memory. The
data is sent in the following order:
1) stx
2) Lot number E.g. “01” for lot No. 1
3) Channel #1 status
E.g. “1” for ppm logging selected in this lot
“0” for ppm logging not selected in this lot
4) Channel #2 status
E.g. “1” for % logging selected in this lot
“0” for % logging not selected in this lot
5) Channel #3 status
E.g. “1” for not used “0” for not used
6) Channel #4 status
E.g. “1” for temperature logging selected in this lot
“0” for temperature logging not selected in this lot
7) begin sample time, min. E.g. “59” for 59 minute
8) begin sample time, hour E.g. “12” for 12 hour
9) begin sample time, day E.g. “09” for 9th day
10) begin sample time, month E.g. “09” for September
11) begin sample time, year
E.g. “96” for year 1996
12) logging interval
E.g.“0” for 1 second
“1” for 15 seconds
“2” for 30 seconds
“3” for 1 minu
“4” for 5 minutes
“5” for 30 minutes
“6” for 60 minutes
“7” for 120 minutes
“8” for 180 minutes
13) Total number of samples per lot
E.g. “1234” for total no. of samples: 1234.
14) Logged data in signed integer, repeat sending in logged
channel sequence