User Guide
18 23
• Salinity Compensation
• Altitude Compensation
• Log Interval
• Time
• Date
• Baud Rate
• Command Prefix
• Buzzer Status
If you press the SETUP key anytime in theView Settings Mode the device will
enter the measurement mode.
CHANGE SETTINGS
To modify a parameter from the setup menu press the CAL key. You will enter
the Change Settings Mode.
If you don’t want to change the parameter press the CAL key. The setting
value will revert to the previously memorized value and the device will return
to the View Settings.
Setting the Salinity Compensation
In the View Settings Mode, press the UP or DOWN arrow keys to display the
salinity factor.
Press CAL to enter the change setting mode, the salinity factor will start
blinking.
Use the UP and DOWN arrow keys to set the
salinity between 0 and 40 g/l.
Press the CFM key to confirm the salinity factor.
Press the CFM key, the day starts blinking.
Use the UP or DOWN arrow keys to select the day.
Press the CFM key and the year on the secondary LCD will blink.
Use the UP or DOWN arrow keys to select the year.
Press the CFM key to confirm the date setting.
Setting the Baud Rate
In the View Settings Mode, press the UP or DOWN arrow keys to display the
baud rate.
The transmission speed (baud rate) of your HI2400 and of the external device
must be the same.
Press the CAL key to enter the change setting mode, the baud rate will start
blinking.